NTU vs FNU: which turbidity unit should a drinking water lab use?

Set up a bench nephelometer or validate a new meter against your SOP, and the unit question answers itself: your approved compliance method already specifies it. For most US drinking water utilities reporting under the Safe Drinking Water Act, that method specifies NTU. NTU and FNU aren't interchangeable, and the real reason has everything to do with how the two methods shine light through water.
How NTU and FNU measure differently
Both units get calibrated against the same formazin polymer standard, which is why the names look almost identical and trip people up. The split happens at the light source and detector geometry.
A nephelometric turbidity unit (NTU) comes from EPA Method 180.1, which specifies a tungsten-filament lamp, a broadband white light source, measured at 90 degrees from the incident beam. A formazin nephelometric unit (FNU) comes from ISO 7027, which specifies a near-infrared light source, usually an 860 nm LED, also measured at 90 degrees. Same angle, same formazin calibration curve, different wavelength, and a different lamp that ages and drifts differently over time.
That wavelength difference matters more than it sounds. Natural particulates (algae cells, clay fines, organic floc) scatter white light and infrared light at different intensities depending on particle size and color. In very clean, low-turbidity source water the two readings often sit close together. Once you've got algal blooms, iron floc, or colored dissolved organic matter in the sample, NTU and FNU can diverge enough to matter for a treatment decision.
Why your compliance method dictates the unit
This is the part that governs what goes on your SOP. EPA Method 180.1 is written around NTU and a specific white-light instrument geometry. Your method approval decides whether an instrument reading in FNU is acceptable as an alternate test procedure for SDWA compliance reporting, or whether it needs a method conversion documented and signed off by your state primacy agency.
A lot of modern benchtop and process turbidimeters display both NTU and FNU on the same screen, sometimes switchable with a menu setting. That convenience has caused real mix-ups on lab bench sheets: an analyst copies the FNU number into a field labeled NTU because the meter showed both and nobody checked which output matched the approved method. Worth a five-minute audit of your current SOP and instrument config if you haven't checked in a while.
Turbidity unit conversion: there isn't a clean one
People ask for a conversion factor between NTU and FNU the same way they'd ask for one between Celsius and Fahrenheit. It doesn't exist, at least not as a fixed multiplier you can apply across the board. Because the divergence depends on particle size, shape, and color in the specific sample, a factor that works for your reservoir in July won't hold for the same reservoir during spring runoff. The only reliable way to compare the two is a side-by-side split-sample run on both instrument geometries, logged over a few seasons so you can see where and how far your source water's NTU and FNU readings pull apart.
Satellite turbidity vs lab turbidity: two different jobs
Satellite-derived turbidity works on a different principle from both NTU and FNU. A multispectral sensor retrieves turbidity, chlorophyll-a, and a clarity estimate from reflectance across a whole lake surface, calibrated to its own reference curve. It won't replace your compliance-grade bench reading at the intake. What it's good for is telling you, before your sampling crew rolls out, which of the lakes in your source-water portfolio need a boat this week and which ones look stable. A weekly read across every lake in your region is built for exactly that triage step, ahead of the grab samples that still go to the bench for the number that counts.
If your monitoring list runs past the one or two lakes you can afford to sample by boat each week, it's worth seeing what Water Quality Retrieval's weekly memo looks like for your region.